1. The Ventilator adopts microcomputer control with high-definition TFT color LCD display, integrating the functions of respiratory control and displaying the monitoring parameters.
It is versatile in function and excellent in performance.
2. The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
3. High-definition four-tube flowmeter, designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
4. Reliable pneumatic transmission system.
5. Economic and proper configuration of technologies, optional as required by the Customer.
6. Modularized design and upgrading. Upgrading is available as required by the Customer.
Part of Anesthesia Machine
7. Main body:
High-strength engineering plastic rack, light, beautiful and corrosion resistant
8. Scope of application: Adult, child
9. Gas source
O2: 0.27-0.55MPa N2O: 0.27-0.55MPa
10. Flow meter
O2: 0.1-1.0L/min 1.1-10L/min
N20: 0.1-1.0L/min 1.1-10L/min
11. O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
12. Flow rate of rapid oxygen supply: 25-75L/min
13. Low oxygen pressure alarm:
There will be sound alarm when the oxygen pressure < 0.2MPa
14. Vaporizer
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
15. Respiratory circuit
Working mode: All-close, semi-close, semi-open
Releasing pressure: 0.5-7 kPa
Respiratory Bellows
Bellows for adults, bellows for children
Tidal volume range: 0-1, 500 ml
Part of Ventilator
16. Display mode: High-definition TFT color LCD screen display
17. Ventilation mode
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV)
18. Ventilation parameters
Tidal volume: 50-1500 ml (20-1500 ml)
Rate: 1-99 bpm
SIMV rate: 2-20 bpm
I: E: 2: 1-1: 8 (4: 1-1: 8)
Inspiratory trigger pressure -1.0-2.0Kpa
PEEP: 0-2.0Kpa
Pressure Range: 1.0-6.0Kpa
SIGH: 1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform 0-50%
19. Parameters for ventilation monitoring
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow-volume oop, pressure-volume loop)
20. Monitoring of oxygen concentration: 21%-100%
21. Security Alarm System
1) Oxygen Concentration alarm: Upper limit setting range 21%-100%
2) low limit setting range: 10%-80%
3) Airway pressure alarm: Upper limit setting range 0.3-6.0 kPa
4) low limit setting range: 0.2-5.0 kPa
5) Per-minute ventilation volume alarm: Upper limit setting range 3.0-30L/min
6) low limit setting range: 1.0-10L/min
7) Sustained high-pressure alarm: It will give alarm when stress have consistently been higher than 2.5 kPa
8) Suffocation alarm: It will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
Power AC 220V± 10% 50Hz± 1Hz (UPS, Storage battery)
22. (Mechanical Arms)
23. (End-expiratory CO2 monitoring)